CONTEC MS200 NIBP Simulator Blood Pressure Monitor Accuracy Simulation Test

CONTEC MS200 NIBP Simulator Blood Pressure Monitor Accuracy Simulation Test GZ Industrial Supplies is the supplier of Contec brand of medical equipment in Nigeria. Our Contec MS200 NIBP Simulator Blood Pressure Monitor Accuracy Simulation Test is a multi-purpose tester applied to the oscillometric Non-Invasive Blood Pressure (NIBP)Monitor. The Contec MS200 NIBP Simulator Blood Pressure Monitor Accuracy Simulation Test can test the performance of multiform blood pressure monitor, provides dynamic blood pressure simulations, static pressure calibration, automated leak testing and pressure relief valve testing.
SKU: RD1029
Call for pricing

Main features of CONTEC MS200 NIBP Simulator Blood Pressure Monitor Accuracy Simulation Test

◆ 4.3" LCD display,interface is concision,operation is simple and convenient.

◆ Provides pressure gauge testing,pressure leak testing,relief valve testing functions .

◆ Build-in pump provides pressure source with leak and pressure relief tests

◆ NIBP simulations for both arm and wrist cuffs.

◆ Provides adult,neonate, seven variations patients conditions,user-defined simulations.

◆ Provides arrhythmias,respiratory artifact simulations.

◆ Built-in air chamber to simulate adult and neonate cuff.

◆ Heart Rate for NIBP Simulations

  • Heart Rate Accuracy:   ±1BPM
  • Patient Conditions:
  • Weak pulse,Tachycardia,Obese ,Geriatric: ±1% ±1BPM
  • Mild exercise:     ±1.5% ±1BPM
  • Strenuous exercise:    ±3% ±1BPM

◆ Pressure measurement

  • Units:      mmHg,kPa,cmH2O,inH2O,psi
  • Range:      0 mmHg-400 mmHg
  • Resolution: 1 mmHg,  0.1 kPa, 1 cmH2O,  1 inH2O,  0.1 psi
  • Accuracy:   0 - 300 mmHg: ±0.5 % of reading ±1 mmHg, 301 - 400 mmHg: ±2 % of reading

◆ Pressure Generation

  • Pressure Generator, Static Pressure Range: 50 mmHg-400 mmHg
  • Difference between target pressure and actual pressure : ±10 mmHg from 100-400 mmHg with a minimum volume of 300 cc
  • Internal leak rate: <2 mmHg per minute, with a minimum volume of 300 cc

◆ Power: 100 - 240 V  AC 50/60 Hz, 60 VA

◆ Environmental Conditions

  • Working temperature:  15℃ to 40℃(59℉ to 104℉)
  • Storage temperature:  -20℃ to +65℃(-4℉ to +149℉)
  • Relative humidity:    < 90%

Specification of CONTEC MS200 NIBP Simulator Blood Pressure Monitor Accuracy Simulation Test

◆ Display: 65k colors 4.3 inch TFT Display, brightness can be changed.

◆ Size : (L x W x H)27 cm x 26.5 cm x 13.2 cm

◆ Weight: 3.5kg